Pots will change the display and you can set up to 2500V and 250mA and 100W respectively, but the High If the remote plug is OK but both sets of output leads are not plugged in, the Voltage, Current and Power None of the control pots will change the display.ĥ. If the remote plug is not correct, the supply will just display 0.00 on the LED display and there will be no Ports (A & B on front panel), or the power supply won't work.Ĥ. You need to have BOTH sets of High Voltage leads (cables) inserted into the High Voltage Output

If you don't have the 'Interlock' connection link on the DB15 connector, the power supply won't work.ģ. If you don't have the remote plug inserted into the blue DB15 connector near the transformer, the powerĢ. In most cases, the units are not faulty but are simply defeated by their optical or electrical interlocks.ġ. I have been contacted by many people asking for advise on how to get their supplies working, because often they often appear faulty and will only display 0 on the red LED output and produce no output voltage. These supplies have been appearing on eBay and there is very little technical information available on the Net. There are a number of safety features incorporated in the supply including optical safety interlocks on the outputsĪnd a remote interlock switch capable of disabling the supply when an enclosure is opened. The power supply produces LETHAL Voltages and must be treated with appropriate care. NEC Pasolink Conversion to 47 Ghz PLL Operation.Testing of Unidentified Radioactive Source.Jabra A210 Bluetooth Interface for Ham Radio.Motorola Spectra (Range 1) UHF VCO Modification for 440 Mhz.